Versatile
standard Capillary and Packed Gas Chromatograph
Get the high
performance of the GC-2010 now also for your routine work. Based on the
successful design of the GC-2010, the GC-2014 gives best results for routine
analysis for a good price-performance ratio. Whether you work with packed or
capillary columns, the GC-2014 provides excellent performance.
Performance
lifted to a higher level
The GC-2014
Series is equipped with the advanced flow controller (AFC) technology inherited
from the GC-17A Series and GC-2010 Series. This digital control is standard for
not only capillary columns, but packed columns as well. Accurate flow rate
control via AFC has higher-level repeatability of retention time and peak area,
enabling a higher level of analyses.
The detectors
have been completely redesigned, incorporating the GC-2010 detector designs for
capillary analyses and the GC-14 detector designs for packed columns. This
TCD-2014 unit is ideal for packed column measurements employing the
semi-diffusion cell design used in TCD-14. The newly designed FPD-2014 takes
advantage of the holophotal construction of the FPD-2010.
Easier to
understand, simpler operation
A large LCD
displays chromatograms and method parameters. Graphical popup screen that
clearly guides operation so that operation errors are prevented. The built-in
Help function almost eliminates need for familiarization training.
Self-diagnostics
validate that the instrument before injection. Regular diagnosis prevents
unexpected downtime.
Expandability
to Support all Types of Analysis
Column replacement is easier in large column oven
|
Select from
four types of injection units and five types of detectors according to the
target compounds and analysis objective. Modular injection units, detectors and
auto-injectors can be easily added after installing the unit in your lab. Three
Injection Units and Four Detectors can be simultaneously installed.
Effective use
of valuable column resources is maintained by using the 40-mm pitch glass
column the same as Shimadzu GC-7, 9, 12, 14, 15 and 16 Series.
A Full Line of
Injection Units and Compact, High-Sensitivity Detectors
Injection Units
DINJ-2014
Dual Packed
Injection Unit
Designed for
Dual FID and TCD analyses. Because two flow paths are handled
using one
temperature control port, these count as one heated zone.
SPL-2014
Split/Splitless Injection Unit
Standard unit
for high-speed analysis with a narrow bore column. The gas saver function
restricts the total gas used. High-pressure injection standard
SINJ-2014
Single Packed Injection Unit
This is a specialized
sample injection unit for use with ECD or other high sensitivity detectors.
WBI-2014 Direct
Injection Unit
Incorporates a
septum purge flow path to restrict solvent tailing. Sharing glass inserts with
splitless analysis simplifies parts requirements (patented).
Detectors
FID-2014
Flame
Ionization Detector
Automatic
ignition and re-ignition functions are standard. By mounting an APC or solenoid
valve unit, a feedback function cuts off gas supply when the hydrogen flame is
extinguished. Generally used for detection of organic compounds with a hydrogen
carbon bond. Single and dual FID's arele. availa
TCD-2014
Thermal
Conductivity Detector
The TCD-2014 is
a semi-diffusion type cell reducing contamination and increasing cell
lifetimes. Incorporates an automatic filament protection circuit. The TCD-2014
is used for analysis of inorganic gases and concentrated organic compounds.
ECD-2014
Electron
Capture Detector
This cell is
very similar to the ECD-14; so spare radiation sources can be shared. This
detector is used for analysis of electrophillic compounds. Improved cell
insulation and reduces contamination achieving higher sensitivity.
FPD-2014
Flame
Photometric Detector
The FPD-2010,
the nozzle system was updated to provide support for packed column analysis
while maintaining its high sensitivity. The FPD-2014 is compact with a high
maximum temperature (350deg.C). This detector is used for analysis of organic sulphur
compounds and organic phosphorus compounds such as residual pesticides and
malodorous components.
FTD-2014
Flame
Thermionic Detector
Packed column
analysis is supported by the FTD-2014 using the FTD-14 mechanism and collectors
from both the FTD-14 and -17. Used for analysis of organic nitrogen compounds
and organic phosphorus compounds such as residual pesticides.
FTD-2014C
Flame
Thermionic Detector
This
specialized capillary-type detector employs the FTD-2010 mechanism. Used for
analysis of organic nitrogen compounds and organic phosphorus compounds such as
residual pesticides. The new collector construction allows replacement without
tools. An optional alkali source regeneration kit recoats the bead reducing
running costs.